Site Logo
language
  • 中文
  • English
  • 日本語
Dealer Log In
Forget PSD
  • About Us
    • Company Profile
    • Kingstate Vision
    • Company History
    • CSR
    • Benefits
    • News
  • Product
    • Murata Cross Item
      Piezo Buzzer
      Magnetic Buzzer
      Speaker
      Dynamic Receiver
      Microphone
      Speaker Module/ Mic Module
      Audio Products
      IP Audio System
  • Capability
    • Total Solution
    • Manufacturing
    • Quality Assurance
    • Technology
    • RoHS / REACH
  • Acoustics
      • Speaker

      • Magnetic Circuit
      • Driver study
      • Polar plot
      • Vibration analysis
      • Sound reflections
      • Box Designs
      • more
      • Headphone Driver

      • Pressure Acoustic
      • VoIP Phone
      • Optimizing Leakages
      • Microphone

      • Discover the Secrets of Sound: ECM Hi-SNR Microphones
      • Receiving Path Fr Simulation
      • Rubber Holder Design
      • Mic Airtight Structure Design
      • Hands-Free Car Mic Module Design
      • Uni-directional Mic Structure Design
      • more
      • Buzzer

      • Resonance frequency simulation
      • The Structure of the Soundhole
      • Side Soundhole Design
      • Waterproof Buzzer
      • Buzzers for Smoke Alarm
      • Multi-tone and Medical Buzzer
  • Applications
    • Security & Smart Home
    • 5G & Smart City
    • Automotive
    • Consumer
    • NB & Tablet
    • VoIP
    • Home Appliance
    • Medical
    • Equipments
  • Contact
RFQ

Inquiry List

Image model Operation
Inquiry Now
  1. Home
  2. News

News

Technical Article

  • Introduction to IEC60601-1-8 Regulations and Medical Product Applications


    The purpose of the IEC60601-1-8 standard is to define the safety and performance requirements for audible or visual alarm systems used in medical electrical equipment. This article only uses the sound alarm as the main description. This standard specifies minimum sound pressure limits and duration for sirens. It is required that the alarm volume be heard in a noisy environment, and the alarm time must be long enough to attract the attention of medical personnel to ensure that correct responses and judgments can be made promptly without causing confusion, misjudgment, or disturbing the patient. Suffer. Therefore, if the alarms in medical equipment can comply with this standard, it will be a guarantee for the medical staff and patients in the hospital.

    【 Main requirements of IEC 60601-1-8 standard 】
    The alarm's pulse frequency must be between 150Hz and 1KHz and the pulse signal must be a square wave. Suppose the alarm is a speaker (usually using a sine wave as the signal input). In that case, it must be ensured that the speaker maintains sufficient power to keep the pulse signal conditions.

    Figure 1, the pulse signal defined in IEC 60601-1-8

    In addition, the speaker's resonant frequency ( fo ) must not be higher than 400Hz to meet the frequency requirement that the pulse frequency between 300Hz and 4 KHz has at least four harmonics (that is, the highest odd number is the ninth harmonic). Since the available space of the speaker equipment will directly affect the sound pressure (in-box resonance), if relevant information such as the size of the space reserved for the speaker in the equipment can be provided, products with suitable specifications can be found more effectively.

    When choosing a speaker as a medical alarm, you should choose a model with a relatively flat frequency response curve to better meet the sound pressure levels at the four resonant frequency points required by regulations ( the sound pressure level error value is within ± 15d B ).

    Figure 2, as smooth as possible frequency response

    【 Applications】

    Table 1 The products applicable IEC60601-1-8

    The above table shows K INGSTATE's speaker products that comply with the IEC60601-1-8 standard. They are widely used in medical equipment such as medical devices, ventilators, blood pressure machines, and medical mattresses. Customers can choose the appropriate model according to the size or form of the medical product they require. K INGSTATE also provides simulation systems that match speakers and products to assist customers in design and development. 


About Us

  • Company Profile
  • Kingstate Vision
  • Company History
  • CSR
  • Benefits
  • News
    • 全部資訊46
    • Environmental1
    • Exhibition News4
    • News18
    • Shareholder news5
    • Technical Article18

10F, No. 69-11, Sec.2, JhongJheng E. Rd.,Danshui Dist, New Taipei City 251402, Taiwan

+886-2-2809-5651

+886-2-2809-7151

info@kingstate.com.tw

志豐電子 Kingstate Electronics Corp.

Kingstate is a well-known and leading acoustic manufacturer in Asia. We are dedicated to providing customers with high-quality products and professional sound solutions.

志豐地圖 Company Map

© 2019 Kingstate Electronics Corp. POWER by OZCHAMP